Operation

21

4.0 Operation

Before connecting any components, follow the

safety information in the ETS-Lindgren

Product Information Bulletin included with your

shipment.

Do not use without spacer cone in place.

About Zeroing

Zeroing determines the offset value that will be subtracted from all subsequent

measurements.

RF must not be present in the vicinity of the probe when the zero

switch is activated, or it will result in an inaccurate zero.

There are three ways to zero a probe connected to the HI-1710A Microwave

Measurement System:

Press the ZERO button on the front panel.

Send the serial port command Z.

Attach an external device to the ZERO connector on the back panel.

The zero will drift as the electronic circuitry and diodes warm up to normal

operating temperature. There will also be small variations as the ambient

temperature changes during the day. For these reasons, the zero point must be

reset periodically.

In addition, the bias and offset readings are checked against preset internal limits

each time the zero switch is activated. If the limits are exceeded, an error

message is displayed. Zeroing in a sufficiently high RF field or zeroing with the

probe removed will also cause an error. These tests measure the performance

characteristics of the probe and serve as failure indicators.