Stim, Reset, E 44 – ETS-Lindgren HI-1710A Microwave Oven Survey Meter User Manual
Page 44
Advertising

44
Menu Options, Settings, Command Set
www.ets-lindgren.com
STIM
Performs a stimulus routine to verify that the probe is operational.
Menu Option
Setting
Serial Port
Command
NO
—
YES
S6
When activated, the HI-1710A will apply a DC voltage to the input of
the RF signal.
The STIM value should be between 2.44 and 4.88. Readings outside
this range indicate a fault in the HI-2623 probe/preamplifier.
RESET
This will reset the HI-1710A to factory settings and overwrite any
settings you changed. See Factory Default Settings on page 46 for
more information.
In addition to resetting the system to factory defaults, RESET performs
the self-test sequence.
Menu Option
Setting
Serial Port
Command
NO
—
YES