Stim, Reset, E 44 – ETS-Lindgren HI-1710A Microwave Oven Survey Meter User Manual

Page 44

Advertising
background image

44

Menu Options, Settings, Command Set

www.ets-lindgren.com

STIM

Performs a stimulus routine to verify that the probe is operational.

Menu Option

Setting

Serial Port

Command

NO

YES

S6

When activated, the HI-1710A will apply a DC voltage to the input of

the RF signal.

The STIM value should be between 2.44 and 4.88. Readings outside

this range indicate a fault in the HI-2623 probe/preamplifier.

RESET

This will reset the HI-1710A to factory settings and overwrite any

settings you changed. See Factory Default Settings on page 46 for

more information.

In addition to resetting the system to factory defaults, RESET performs

the self-test sequence.

Menu Option

Setting

Serial Port

Command

NO

YES

I