IDS 430/433 Users Manual Version 1.0

- 13 -

Setup Guide

Initialize to Factory Defaults

Setup the IDS 433/430 in the following order:
The IDS 433/430 is initialized at the factory. The IDS 433/430 should be

initialized when first powered up to guarantee no corruption occurred during

shipping. It must be re-initialized if the program memory (EPROM) or the

parameter memory (EAROM) is replaced. The Initialize function clears the

memory and sets the configuration parameters to their default values. Use

parameter function 70 described in

Parameter Functions

section (Page 7-7) to

initialize the IDS 433/430.

NOTE: Use the code 10809 for a factory default initialization. Calibration

and setup must be performed after the IDS 433/430 is initialized with this

code.

Setup & "Quick" Calibration Guide

The IDS 433/430 should be configured after initializing and

BEFORE

calibration.

Set Operation Parameters

details all of the scale parameters that are available.

To quickly get the IDS 433/430 up and running, the following Quick Calibration

Guide lists only the mandatory parameters that must be defined. Enter the

following scale parameters using the selection outlined in

Set Operation

Parameters

and

Parameter Functions

.

Parameter (section)

Purpose (Factory Setting)

1 (Set Operation Parameters)

Load Cell mv/V (3)

11 (Set Operation Parameters)

Decimal Point (none)

12 (Set Operation Parameters)

Count-by (1)

17 (Set Operation Parameters)

Full Scale Graduations

(10,000)

18 (Set Operation Parameters)

Overload Graduations

(10,200)

60 (Parameter Functions)

Calibrate Scale

Calibrate the Weight Indicator

Use Parameter Function 60 to calibrate the weight indicator (page 7-2).