Industrial Data Systems IDS 430 User Manual

Page 45

Advertising
background image

IDS 430/433 Users Manual Version 1.0

- 45 -

Function 64 Configure Passwords

Use function 64 to enter password data. The IDS 430/433 provides three

security passwords that can be activated and entered to protect configuration

data. Password 1 is used to protect I/O port configuration, setpoint, and event

monitor data. Password 2 is used to protect calibration and scale configuration

data. Security level 3 provides extra protection for password level 2 data.

Security level 3 requires the entry of password 2 and the Function Lock inside

the IDS 430/433 to be in the unlocked position. Password 3 will inhibit viewing,

changing, and uploading of Scale Basic functions to a PC when password is

active but will allow downloading a new Scale Basic program from the PC to the

indicator. All passwords will be disabled once the new program has been

downloaded into the indicator and should be reactivated.
The Operation Parameter Reference Table lists the configuration parameters and

the password/security level required to access them. The IDS 430/433 prompts

“PASS 1” when a parameter is selected that requires password level 1. If the

entered password is incorrect, the IDS 430/433 displays “Error”. Press the

ENTER key to try again. Press the CLEAR to exit from password entry.

IMPORTANT NOTE

: Write down the passwords and put them in a safe place. If

you loose and forget a password, there is no way to access the protected

functions.
1. Hold the CLEAR key down and press the ENTER key to enter the configure

mode.

2. Enter CFG 64. The IDS 430/433 prompts “P1 xxxx” where xxxx is the current

password number. Enter a new password number or press the ENTER key

to use the displayed number. Enter a 0 (zero) for the password number to

disable the password.

3. The IDS 430/433 prompts “P2 xxxx”. Enter password 2 data.

4.

The IDS 430/433 prompts “P3 xxxx”. Enter password 3 data.