Industrial Data Systems IDS 430 User Manual

Page 55

Advertising
background image

IDS 430/433 Users Manual Version 1.0

- 55 -

Function 74 Configure Timers

The display prompts “tr” when function 74 is activated. Enter the timer number to

be modified (1 to 5) then press the ENTER key. The first parameter (L) is the

time length in tenths of a second. The second parameter (Fn) is the function

number to be executed when the timer times out.

Parameter 75 Enable Peak Detect

Enter a 1 to enable peak detect, enter a 0 to disable peak detect.

Parameter 76 Debug mode for Event monitors and Scale Basic

The Debug mode is used to ‘debug’ event monitors and scale basic functions.

Enter a 1 to activate debug mode, enter a 0 to disable debug mode. When debug

mode is enabled Scale Basic functions ‘single step’. The subroutine stack level

and the Function number are displayed when entering a function. Example: “0

F12”. Press the ENTER key to begin execution of the function. The instruction

address and instruction code are displayed. Example: “21 255”. Continue to

press the ENTER key to step thru the program. NOTE: The first step displayed

for all scale basic functions is a GOTO (244).

Function 78 Initialize ID memory

Use function 78 to initialize id memory after it has been corrupted. The ID

memory becomes corrupted when power is lost (when battery backup is

disabled) or when the battery is allowed to discharge below 2 volts.

Parameter 79 Enable Multi-point Linearization

Enter a to 1 to enable multi-point Linearization, enter a to 0 to disable multi-point

Linearization. The IDS 430/433 uses the calibration factors set in function 80

below when multi-point Linearization is enabled.

Advertising
This manual is related to the following products: