IDS 430/433 Users Manual Version 1.0

- 34 -

Parameter 14. Alternate Decimal Point (1)

Enter the decimal point position for the alternate units display.

Parameter 15. Alternate Count-by (5)

Enter the alternate units count-by.

Parameter 16. Alt Units Conversion Factor (45,360)

The IDS 430/433 calculates the alternate weight by multiplying the primary

weight by the conversion factor and then dividing by 100,000. The following table

lists some common conversions.

Alternate Units Conversion Factor Table

Conversion

Conversion Factor

Conv. Fact. x 100,000

lb to kg

0.45359 kg/lb

45,359

kg to lb

2.20462 lb/kg

220,462

lb to ton

0.0005 Ton/lb

50

kg to tonne

0.001 tonne/kg

100

gram to oz

0.352739 oz/gram

352,739

oz to gram

28.3495 gram/oz

2,834,950

lb to lb

kb to kg

Etc.

1

100,000


Enter the conversion factor x 100,000 for the configured alternate units.

Parameter 17. Full Scale Graduations (10,000)

Enter the number of graduations at full scale. The full-scale graduations are

calculated using the count-by, decimal position, and full-scale weight.


Full Scale Graduations = Full Scale Weight / count-by.

For example, if Full Scale Weight = 600.0 and count-by = 0.2 then Full Scale

Graduations = 600.0 / 0.2 = 3000.

Parameter 18. Overload Graduations (10,200)

Enter the number of graduations at which the scale is in overload status. The IDS

430/433 displays “Err oL” on the numeric display when the scale weight exceeds

overload. For example, if the Full Scale Graduations is 10,000 grads and the

scale is rated for 2% overload, then Overload Graduations = 10,000 + (10,000 x

0.02) = 10,200.

Parameter 19. Deadload Offset

This parameter is automatically calculated by the configure function (60 & 80).