Function 69 Diagnostic Tests

1. Hold the CLEAR key down and press the ENTER key to enter the configure

mode.

2. Enter CFG 69. The IDS 430/433 prompts “diA xx” where xx is the currently

selected test number. Enter a test number or press the ENTER key to use the

displayed number.

3. The IDS 430/433 begins the selected test. Press the CLEAR key when the

test is complete.

Test #

Hardware Tested

Description

1

Serial Com. Port 1

Display input data.

2

Serial Com. Port 1

Display input error count.

3

Serial Com. Port 1

Transmit data.

4

Serial Com. Port 2

Display input data.

5

Serial Com. Port 2

Display input error count.

6

Serial Com. Port 2

Transmit data.

7

Parallel Port

Transmit data.

8

A/D converter

Display raw conversion data.

9

Ram Memory

Test memory for errors.

10

Display

Turn on all LED’s and display segments.

11

Configuration

Print the EAROM configuration data.

12

Serial Com. Port 1

Transmit/Receive loop-back test.

13

Serial Com. Port 2

Transmit/Receive loop-back test.

See section: Diagnostic Tests for more information on each test.

Function 70 Initialize Operation Parameters to Factory Defaults

Use function 70 to reset the IDS 430/433’s operation parameters to their factory

default values. This function is used to reset the IDS 430/433’s earom when a

new EPROM program is installed. Do NOT use this function unless you are

ready to begin installation from the beginning.
1. Hold the CLEAR key down and press the ENTER key to enter the configure

mode.

2. Enter CFG 70. The IDS 430/433 prompts, “rESEt”. Press the ENTER key to

reset operation parameters or press the CLEAR key to exit without changing

the operation parameters.

3.

NOTE

: enter 10809 when the display prompts, “rESEt” to reset the

calibration factors to the factory defaults.