Function 80 Calibrate - Multi-point Linearization

Use function 80 to calibrate the scale using up to 5 calibration points to 'linearize'

the load cell. Set the Scale Operation Parameters before using function 80. Best

results are obtained when the IDS 430/433 and load cells have reached thermal

equilibrium (power applied for approximately 30 to 120 minutes depending on

load cell size). NOTE: function 80 automatically adjusts parameter 19 (deadload

offset).
1. Hold the CLEAR key down and press the ENTER key to enter the configure

mode.

2. Enter CFG 80. The IDS 430/433 prompts "dEAdld". Remove all weight from

the scale. Press the ENTER key to obtain the deadload weight. Press the

Units key (

) to continue without reading a new deadload weight.

3. The IDS 430/433 ‘s display alternates between "CAL 1" and the weight on

the scale. Place the first calibration weight on the scale. Enter the calibration

weight using the numeric keypad. Press the ENTER key. The IDS 430/433

reads the calibration weight.

4. The IDS 430/433 ‘s display alternates between "CAL 2" and the weight on

the scale. Repeat step 3. The IDS 430/433 will accept up to 5 calibration

points. Press the CLEAR key to complete calibration if less than 5 calibration

points are used.

5. The IDS 430/433 prompts "CFG 80". Press the CLEAR key to exit the

configure mode.