IDS 430/433 Users Manual Version 1.0

- 18 -

# 1 Fill to Setpoint

Press the F1 key, the display prompts “SEt Pt” key in the setpoint value and

depress the Enter key. Place an empty container on the scale. Press the F2 key

to begin filling. TTL output 1 turns on until the weight on the scale is equal or

greater than the setpoint.
Connect TTL out put 1 to a relay to turn on motors, open and close gates, turn

on and off flow valves for automated filling or indication.

Installation

Connect TTL output 1 to relay or lamp actuator.

TB4 - Pin 1 = TTL output 1 (active low) +5volts idle, 0 volts active.

TB3 - Pin 1 = +5 volts

Setup Parameters:

(See chapter 6: Set Operation Parameters for setup details)

Parameter 43

Minimum scale weight to begin filling. Verifies that a container is

on the scale. Defaults to 0 = there is no minimum.

Parameter 44

Minimum fill amount. Defaults to 0. Set to some small value to

insure that a fill amount has been entered.

Parameter 45

Maximum fill amount. Default 5000. Maximum fill weight that will

ever be used.

Parameter 46

Scale empty weight. Default 50. Maximum weight to begin fill.

Verifies that an empty container is on the scale.

Operators Functions

Enter setpoint

Press

F1

key to enter the setpoint. Enter the setpoint weight, then press the

ENTER key.

Start fill operation

Place the container on the scale (if required). Press

F2

key to begin fill

operation.

Stop fill operations

Press the

Clear

key to stop fill operation.

Print Results

Press the Print key to print the Gross, Tare, and Net results of the fill operation.

Error Codes

Error 30 = Scale not empty

Error 31 = Minimum weight error: container is not on scale

Error 32 = Tare weight error: tare function failed, scale not stable (motion)

Error 33 = Setpoint less than minimum setpoint amount

Error 34 = Setpoint more than maximum setpoint amount