6. Testing the instrument

Due to the underlying measurement principle,

no calibration of the instrument is required. It is

recommended, however, to check the functionality

of the instrument at regular intervals (about once

every 3 months). The reference tile included with

delivery is provided for this purpose.

Select the “Standard” in the Measure menu. The

configuration of this organizer is fixed and cannot

be changed.

The unit then switches to the measurement mode.
Place the instrument on the test tile and perform

three measurements.
The values printed on the reference tile are the

corrected readings. If the values measured on

the reference tile are within the printed tolerance

range, the requirements are met.
If the mean value is not within the desired

tolerance range, try carefully cleaning the test tile

(see Chap. 16). If this produces no improvement,

please contact our Customer Service department.
Reference tile
To ensure a precise instrument test, only original

test tiles from the manufacturer should be used.

Their surface must not be touched and must be

protected against scratches. Due to environmental

influences, however, the values of test tiles can

change over the course of time even if they are

handled gently. For this reason, have the test tiles

checked by the manufacturer at regular intervals

(annual checks are recommended).