Info and error messages
Identifies an internal error. Load a firmware
update. When in doubt, call the Customer Service
department.
The measurement unit was moved across the
sample too fast. The measurement is invalid and
must be repeated.
The instrument was accelerated too fast across the
specimen or the scan direction was changed too
fast.
Confirm the message by pressing “operate” and
repeat the measurement.
Appears when the “Plausibility Control“ function in
the Configuration of the used Memory or Organizer
file is ON and the corrected value deviates by more
than 20% from the uncorrected value. Repeat the
measurement.
Not enough light is reflected by the surface.
Repeat the measurement on an appropriate
sample area.
The operate key was released before the
full scanlength was reached. Perform a new
measurement.
Day or month falls outside the valid range of 1 - 31
or 1 - 12. Repeat the entry.