ALTANA Micro-Wave-Scan Manual User Manual

Page 53

Advertising
background image

52

Info and error messages

Identifies an internal error. Load a firmware

update. When in doubt, call the Customer Service

department.

The measurement unit was moved across the

sample too fast. The measurement is invalid and

must be repeated.

The instrument was accelerated too fast across the

specimen or the scan direction was changed too

fast.

Confirm the message by pressing “operate” and

repeat the measurement.

Appears when the “Plausibility Control“ function in

the Configuration of the used Memory or Organizer

file is ON and the corrected value deviates by more

than 20% from the uncorrected value. Repeat the

measurement.

Not enough light is reflected by the surface.

Repeat the measurement on an appropriate

sample area.

The operate key was released before the

full scanlength was reached. Perform a new

measurement.

Day or month falls outside the valid range of 1 - 31

or 1 - 12. Repeat the entry.

Error Messages

Advertising