9.1 Number of Measurements

This function allows selection of the numbers of

readings to be taken per sample. If the number is

greater than 1, the measurements are statistically

evaluated. Turning the scroll wheel adjusts the

number while pressing it shifts the decimal place

one further over. Pressing operate completes the

process and a confirmation display appears.

9.2 Scanlength

The setting options for Scanlength are shown in

the neighboring illustration. The current setting is

identified by a black dot at the end of the line.

Move the cursor to the desired setting and press

the operate button. The instrument returns to the

previous menu.
In the setting 0 cm, you can perform

measurements without scan motion. However, you

can only determine the scales du, Wa, Wb and

DOI here.

Note:
Using a short scanlength will result in strong

statistical skewing of measurement values for large

wavelengths. Thus, when using the Wd scale with

a scanlength of 5 cm, at least 3 measurements

should be made per checkzone to ensure

representative values.