1 basics, Etup and, Alibration – Flintec FT-15D User Manual

Page 24: 1 basic setup keys on the frontplate, 2 entering the setup and calibration menu, 3 exiting the setup and calibration menu

Advertising
background image

FT-15(D) Technical Manual, Rev. 1.02 August 2010

Page 24 of 52

9 S

ETUP AND

C

ALIBRATION

9.1 Basics

9.1.1 Basic Setup Keys on the Frontplate

The symbols located on the lower right corner of each key indicate the function of the keys in the setup menu.
The basic meanings of these keys are given in the table below.

Enter

Exit without

saving

Stepping forwards

to the next

parameter

Stepping backwards

to the previous

parameter

Entering the

parameter

value

Changing the

value

Clear (delete)

If you use a PC the Flintec IndFace software can be used for the setup and the calibration of FT-15(D).

9.1.2 Entering the Setup and Calibration Menu

To enter the Setup and Calibration Menu:

Press <F> and <Info> keys sequentially

First [FunC ] and then [PASSWr ] prompts will be displayed sequentially.

Here you can enter the parameter block by using:
the service password:

Press <Zero> key and <Info> key and <Enter> key

or the operator password:

Press <Transit> key and <Enter> key

The calibration jumper (see chapter 4.1) should be short circuit to change the legal metrological parameters in
the service mode (there is black box on the top right corner of the grey coloured boxes in the calibration menu
flowcharts, see Appendix 1).

The Setup and Calibration menu consists of main blocks which are displayed as [X-- ] and sub-blocks.
By using <

> and <> keys you can reach previous or next main blocks. After reaching the desired main block

you can get in by pressing the <Enter> key. As you enter the block you will reach the first sub-block in that main
block. The sub-block address will be seen on the display as [X0- ].
You can also search between the sub-blocks by using <

> and <> keys and reach the first parameter of the

displayed sub-block by pressing the <Enter> key. The number of the parameter is displayed as [XY0 ]. Again
you can search between parameters by <

> and <> keys.

FT-15(D) has numerical keys for entering parameter values.

9.1.3 Exiting the Setup and Calibration Menu

Whatever parameter block you are in – if you press the <F> key, you will get out of the active sub-block and
reach the next sub-block. If you press the <F> key again you will get out of the active block and reach the next
main block. If you press the <F> key once again, the [SAvE ] message appears on the display.
Here you can press

 the <Enter> key to save the changes into the memory or

 you can press the <Info> key to store the changes temporarily until the power goes off or

 you can press the <F> key to abort any changes.

The [Waıt] message will be displayed for a few seconds and afterwards the weighing mode will restart
automatically. Especially for legal metrological usage, please don’t forget to turn the power off and remove the
calibration jumper before you start the operation.