FT-15(D) Technical Manual, Rev. 1.02 August 2010

Page 36 of 52

Transit weighing

11.2.7 Setup

[0--] Interface Block

You can access the parameters of the instruments’s serial interface(s) in parameter section [0--].
The data output modes can be used only once per instrument except for continuous data output.

[00-] Serial Interface 1

The parameters of the 1

st

serial interface (= standard serial interface).

[000 4] Data Format

0 : No data transfer

1 : Continuous data output

2 : PC-Software interface mode (with ASCII P and Z) 3 : Host mode
4 : Ticket and report printer (default)

[001 3] Baud Rate

0 : 1 200 Baud

1 : 2 400 Baud

2 : 4 800 Baud

3 : 9 600 Baud (default)

4 : 19 200 Baud

5 : 38 400 Baud

6 : 57 600 Baud

[002 0] Handshake

0 : No Handshake (default)

1 : Xon/Xoff

[003 00] Address

The address range is 1 to 99. If you enter 0, the indicator will operate without an address.

[004 0] Data Length and Parity

0 : 8 bit, no parity (default)

1 : 7 bit, odd parity

2 : 7 bit, even parity

[005 0] Checksum

0 : Checksum byte disabled (default)

1 : Checksum byte enabled

[01-] Serial Interface 2

[02-] Serial Interface 3

This sub-block includes the parameters of the 2

nd

serial interface respective 3

rd

serial interface.

[010 0] Data Format

respective

[020 0] Data Format

0 : No data transfer

1 : Continuous data output (default)

2 : PC-Software interface mode (with ASCII P and Z) 3 : Host mode
4 : Ticket and report printer (default)

[011 3] Baud Rate

respective

[021 X] Baud Rate

0 : 1 200 Baud

1 : 2 400 Baud

2 : 4 800 Baud

3 : 9 600 Baud (default)

4 : 19 200 Baud

5 : 38 400 Baud

6 : 57 600 Baud

[012 0] Handshake

respective

[022 0] Handshake

0 : No Handshake (default)

1 : Xon/Xoff

2 : Hardware (for serial interface 2 only)

Warning:

If you choose Hardware as handshake, then the serial interface 3 can not be used as RS232C.

[013 00] Address

respective

[023 00] Address

The address range is 1 to 99. If you enter 0, the indicator will operate without an address.

FLINTEC GmbH
Tel : +49 6226 9240 0
Fax : +49 6226 9240 99
Meckesheim, Germany

WEIGHING TICKET

Ticket no

: 22

Date & Time

: 24. 03.2009 09:18


Vehicle ID

: 123456

Customer

: Customer D

Commodity

: 316 L/ 2.5mm


Weight

:

14120 kg