] scale build, 210 0] scale type, If single range has been selected: [212 ] capacity – Flintec FT-15D User Manual
Page 28: Cap1, 212 ] 2. capacity, Cap2
FT-15(D) Technical Manual, Rev. 1.02 August 2010
Page 28 of 52
[208 X] Stability Time (for FT-15D only)
If the scale is in no motion during this time period, the scale is regarded as stable. Allowed settings: 0.0….9.9 s
[21-] Scale Build
The capacity and the resolution of the scale will be defined here.
[210 0] Scale Type
The capacity and resolution parameters will vary according to the selection here.
0 = Single Range (default)
1 = Dual Range
You will reach the next parameter by pressing <Enter>. Please note that the following parameters will vary according
to this selection.
If single range has been selected:
[212 ] Capacity
Press the <Tare> key to reach this parameter.
[CAP ]
[ XXXXXX ]
The capacity of the scale will be entered here by <Tare> and <Zero> keys.
The value will be confirmed by pressing <Enter>.
[d ]
[ XXXXXX ]
Display resolution of the scale will be selected by <Zero> key.
The selection will be confirmed by pressing <Enter>.
If dual range has been selected:
[211 ] 1. Capacity
Press the <Tare> key to reach this parameter.
[CAP1 ]
[ XXXXXX ]
The capacity of the first weighing range (Select from Appendix 5) will be entered here by
<Tare> and <Zero> keys and confirmed with <Enter>.
[d1 ]
[ XXXXXX ]
Display resolution of the first weighing range will be selected by <PV> key and confirmed
with <Enter>.
[212 ] 2. Capacity
Press the <Tare> key to reach this parameter.
[CAP2 ]
[ XXXXXX ]
The capacity of the second weighing range (Select from Appendix 5) will be entered here by
<Tare> and <Zero> keys and confirmed with <Enter>.
[d2 ]
[ XXXXXX ]
Display resolution of the second weighing range will be selected by <Zero> key and
confirmed with <Enter>.
Attention:
After the scale parameters have been changed the scale has to be calibrated.