10 a, ] metrological data block, ] legal metrological records – Flintec FT-15D User Manual

Page 30: Libi, Emory and, Egal, Etrological, Ecords

Advertising
background image

FT-15(D) Technical Manual, Rev. 1.02 August 2010

Page 30 of 52

10 A

LIBI

M

EMORY AND

L

EGAL

M

ETROLOGICAL

R

ECORDS

FT-15(D) keeps the last 9 583 weighing records in the loop alibi memory. The ticket number, which is also the
alibi record number, cannot be changed manually. This data records can be shown on the indicator’s display or
can be transmitted via serial port.
The reports initiated by the report keys located on the instrument’s front can also be used as an alibi memory
report, if they include the required records.
Empty records will be printed as ------ , the characters “PT” are located on the right of a preset vehicle weight
value, the message “Transit“ will be printed in the first weighing column of a transit weighing record and
corrupted records will be printed as ******. For the records, where a weighing result is not available, the
message [ no rec ] will be shown on the display.
After installing the alibi board or changing the main board “Error 41” or “Error 42” will be shown on the display
after power on because the activation of the alibi memory is required.
If the calibration jumper is short circuited, you can activate the alibi board by pressing the <Enter> key after
power on. Otherwise the alibi errors will disappear by pressing the <Enter> key, but you cannot export any data
(Error “40” will be shown after pressing the <Reprint> key ).

Alibi report form:

ALIBI MEMORY RECORD FILE
09.01.2009 17:10

Ticket

Date Time Vehicle

Customer

Commodity

1.Weight kg 2.Weight kg Net kg

---------- ------------- --------- --------------- ------------------ ------------------ --------------- ----------------- ----------

000234

01.01.09 08:12

1111

Customer A

Sample A

33320 11100

22220In

000235

03.01.09 09:42

2222

No entry

No entry

Transit 44440

-------Tr

000236

08.01.09 12:10

3333

23456

123456

11100 PT 55560

44460PV

000237

12.01.09 17:22

4444

Customer B

Sample B

12220 22220 10000Out

000238

13.01.09 09:13

0

0 33330 33330PC


To find a definite record you have to enter parameter [802]. After reaching the desired record, the recorded data
can be viewed in the display. If needed you can print this record together with the following 9 records by
pressing the <Print> key. If you press the <Print> key in parameter [803], all recorded data will be sent via serial
port. The <F> key cancels the print job.

[8--] Metrological Data Block

The parameters about Metrological Registry are being entered in this section.

[80-] Legal Metrological Records

In this block edit the parameters with <

> key and use the <Enter> key according to the parameter’s

description.

[800] Counter

This counter increases automatically by 1 after starting the setup mode with enabled calibration jumper and service
password. This counter cannot be changed manually.

[801] Alibi

Memory)

0 = Disabled

1 = Enabled

[803]

Print all Alibi Memory records

All data from the Alibi Memory can be printed by pressing the <Print> key. You can stop printing with the <F> key.

[804] Alibi

Info

You can get some basic information about the alibi memory board and the records by pressing the <Print> key.
These data are:

 SN :

Pin code of the alibi memory card

 SA :

The alibi record start number within this indicator

 LA :

The alibi record number of the next weighing with this indicator

 CN :

Calibration counter number

 CS :

Check sum status

 AS :

Capacity of the alibi memory

 LD :

Firmware date of the alibi memory

Advertising