3 configuration parameters, ] configuration block [10-] application, ] filter – Flintec FT-15D User Manual

Page 27: ] entries, 4 scale parameters [2--] scale block, ] set up

Advertising
background image

FT-15(D) Technical Manual, Rev. 1.02 August 2010

Page 27 of 52

9.3 Configuration

Parameters

Warning:

The parameters are automatically limited to legal for trade settings if parameter [200] = 1.

[1--] Configuration Block

[10-] Application

In this sub-block you can select the application mode and enter the parameters related to the selected mode.

[100 0] Operation

0 : All Weighing Types (default)

1 : No Preset Vehicle Weighing

2 : No Transit Vehicle Weighing

3 : Only First / Second Weighing.

[12-] Filter

In this block the proper filter values according to the operating conditions can be entered. One of the most
important features of the indicator is viewing the filter characteristics on the display and with the help of this
option, you can select the most suitable filter without leaving the setup mode.

[120 4] Filter

The filter value can be set from 0 to 4. When you enter this parameter and press the <Enter> key while [120 X] is
shown on the display, the weight variation can be seen on the display. The value of the filter can be changed by
using the <

> and <> keys and the weight variation for each value is shown on the display. After finalising the filter

selection you can go to the next step by pressing the <F> key.

[14-] Entries

In this block you can enter date and time.

[140 ] Date

[DD.MM.YY]

First press the <> key for the date entry, then enter the new date via numerical keys and save
by pressing <Enter>.

[141 ] Time

[ HH.MM]

First press the <> key for the time entry, then enter the new date via numerical keys and save
by pressing <Enter>.

9.4 Scale

Parameters

[2--] Scale Block

The parameters related to the weighing operation are located here.

[20-] Set Up

[200 0] Approved

0 = No

1 = Yes

[201 0] Increased Indication (10x)

0 = By pressing key

1 = Always increased indication

[202 2] Power On Zero

This parameter setting controls if after power-on the scale will get automatically zeroed if the weight is in the
percentage of the zeroing range. If the weight is not in the zeroing range the display will show [E E E ] message until
you press the <Enter>.

0 = disable

1 = ± 2%

2 = ± 10%

[203 1] Zeroing Range

0 = disable

1 = ± 2%

2 = ± 20%

[204 0] Auto Zero Tracking

AZT automatically re-adjusts the scale to zero for compensating defined small deviations around the centre of zero.

0 = disable

1 = ± 0,5e

2 = ± 1e

3 = ± 3e

[206 1] Motion Detector

This parameter defines the sensitivity level which will determine what is considered as stable.

0 = ± 0.3e

1= ± 0.5e

2= ± 1e

3= ± 2e

4 = No motion detector