Introduction – YSI ProDSS User Manual

Page 13

Advertising
background image

11

Introduction

Sensor guard and weight installation

1. Carefully slide the sensor guard over the bulkhead and attached sensors/port plugs. Push the sensor guard

toward the bulkhead until the sensor guard threads align with the bulkhead threads.

2. Carefully finger-tighten the sensor guard clockwise.

NOTICE: If any resistance is felt, loosen the sensor guard completely to prevent cross-threading. Incorrect

installation may cause damage to the sensor guard or bulkhead that is not covered by the warranty.

NOTICE: Do not submerge the bulkhead without a sensor or port plug installed in all ports.

Figure 6 Sensor guard and weight installation

1 Sensor guard

4 Depth sensor (if equipped)

2 Bulkhead

5 Weight

3 Bulkhead threads

1

2

3

4

5

Advertising