Maintenance and storage
Conductivity/temperature sensor

NOTICE: Use care when handling the conductivity/temperature sensor to prevent any impact on the exposed

thermistor.

Figure 62 Channel brush

Conductivity/temperature sensor maintenance

1. Dip the sensor’s cleaning brush (included with the maintenance

kit) in clean water.

2. Insert the brush at the top of the channels, and sweep the

channels 15 to 20 times (

Figure 62

).

NOTICE: If deposits have formed on the electrodes, use a mild

solution of dish soap and water to brush the channels.
For heavy deposits, soak the sensor in white vinegar to
assist cleaning, then scrub with the cleaning brush after
soaking.

3. Rinse the channels with clean water following the sweepings or

soak.

Figure 63 Conductivity/Temperature

Short-term storage

Conductivity/temperature sensor short-

term storage (less than 4 weeks)

When in regular field use, the conductivity/temperature sensor should
remain installed on the bulkhead in a dry or water-saturated air
environment.

Place approximately 0.5 in (1 cm) of any water (deionized, distilled or
environmental) in calibration cup.

Install the calibration cup on the bulkhead and firmly tighten to prevent
evaporation (

Figure 63

).