Calibration

Figure 51 Calibrate pH 2- or 3-point

pH calibration 2- or 3-point

NOTE: If performing a 2- or 3-point calibration, one point should

be in buffer 7; however, the calibration points can be in any
order.

1. Perform steps 1-7 of the pH calibration 1-point procedure (

pH

calibration 1-point on page 39

).

2. Rinse the sensor 2-3 times with a small amount of pH 4 or pH 10

buffer solution.

3. Rinse, then fill the calibration cup to the appropriate level with

the buffer solution (approximately 170 mLs) that is the same value
(pH 4 or pH 10) used to rinse the sensor.

4. Carefully immerse the sensors into the solution.
5. Allow at least one minute for temperature stabilization, then

select Calibration value.

6. Enter the pH buffer value that corresponds to the buffer

temperature reading (value may be located on pH buffer bottle).

7. Observe the actual measurement readings for stability (white line

on graph shows no significant change for 40 seconds), then select

Accept Calibration (

Figure 51

). “Ready for cal point 3” will be

displayed in the message area.

8. Select Finish Calibration for a 2-point calibration or repeat the 2-

or 3-point calibration procedure with the third buffer solution.

Figure 52 Calibrate ORP

ORP calibration

1. Obtain/prepare a standard with a known oxidation reduction

potential (ORP) value.

NOTE: YSI recommends Zobell solution.

2. Fill the solution to fill line 1 on the calibration cup (approximately

170 mLs).

3. Carefully immerse the sensors into the solution.

4. Push the

Cal

key, then select pH/ORP, then ORP.

5. Allow the temperature of the standard to stabilize then select

Calibration value.

6. Enter the ORP calibration value that corresponds to the measured

temperature reading (example: the value of YSI Zobell @ 20 �C is
237.5)

7. Observe the actual measurement readings for stability (white line

on graph shows no significant change for 40 seconds), then select

Accept Calibration (

Figure 52

). “Calibration successful!” will be

displayed in the message area.