Maintenance and storage
Depth sensor maintenance and storage

Figure 59 Depth sensor flush

Depth sensor storage

The ProDSS optional depth sensor accesses water through ports located
in the bulkhead (

Figure 59

). Although not directly accessible, correct

maintenance and storage is necessary for reliable operation.

The depth sensor can be stored dry, in water-saturated air or submerged
in water.

NOTICE: To prevent damage to the sensor’s strain gauge, do not

store the sensor in corrosive solutions.

Depth sensor maintenance

Periodically clean the depth ports with the syringe included in the
ProDSS maintenance kit (626990). Fill the syringe with clean water and
gently force water into one of the ports. Flush until clean water flows from
the opposite depth port.

NOTICE: Do not insert objects into the depth ports. Damage

to the depth transducer from incorrect cleaning is not
covered by the warranty.