Operation – YSI ProDSS User Manual

Page 20

Advertising
background image

18

Operation

Figure 13 GLP Options

GLP Options

GLP → Options

User ID, Probe ID, or User Field #1 or 2 can be user-defined for positive
GLP file identification of:

• The person calibrating the instrument.
• The sensor/cable serial number used during calibration (or other,

user-defined Probe ID).

• Other user-specific identification (User Field #1 and #2) (

Figure 13

).

NOTE: User Field can be used to describe the condition of the

probe. For example, new sensor or new ODO cap.

Figure 14 Re-Cal Prompts

Re-Cal Prompts

GLP Options → Re-Cal Prompts

Re-Cal Prompts provide a reminder to recalibrate a probe in the
user-defined number of days (

Figure 14

).

The Re-Cal prompt will be displayed in the message area of the main
display when the set time has elapsed (

Figure 9 on page 15

).

Select the desired sensor Re-Cal prompt, then enter the desired number
of days before the Re-Cal prompt occurs.

Set the sensor value to zero (0) days (default) to turn off Re-Cal prompts.

NOTE: When enabled and the set amount of time since the last

calibration has passed, the Re-Cal prompt will be shown
when the instrument is turned on.

Figure 15 GLP Security

GLP Security

GLP → Security

The Calibration menu can be password protected to prevent accidental
or unauthorized sensor calibration (

Figure 15

).

1. From the GLP menu, select Security, then enter the default

password “ysi123”.

2. Select Set Password [] and change the default password.
3. Select the Protect Cal check box to password protect the

Calibration menu.

NOTE: Write down and keep the password in a safe place. Contact

YSI Technical Support if you lose the password (

Technical

support on page 81

).