Maintenance and storage

Figure 67 ODO rehydration

ODO sensor rehydration

To prevent sensor drift, always store the ODO sensor in a wet or
water-saturated air environment. If the ODO sensor has accidentally
been left dry for longer than 8 hours, it must be rehydrated.

If rehydration is necessary, soak the ODO sensor cap in warm (room
temperature) tap water for approximately 24 hours. After the soak,
calibrate the sensor (

Figure 67

).

Figure 68 ODO short-term storage

ODO sensor short-term storage (less than 4 weeks)

When in regular field use, the ODO sensor should remain installed on
the bulkhead. Place approximately 0.5 in (1 cm) of any water (tap or
environmental) in the calibration cup (

Figure 68

).

Install the calibration cup onto the bulkhead and firmly tighten to prevent
evaporation.

Figure 69 ODO long-term storage

ODO sensor long-term storage

The ODO sensor can be left on the bulkhead or removed for long-term
storage (

Figure 69

).

Installed on bulkhead

Fill the calibration cup with clean water (use distilled or deionized water if
a pH sensor is not installed). Submerge the sensor in the calibration cup
then firmly tighten to prevent evaporation.

Removed from bulkhead

Remove the sensor from the bulkhead (

Sensor removal on page 10

).

Method 1 : Cover the sensor connector end with the plastic storage

cap. Submerge the sensing end of the sensor in a container
of clean water (use distilled or deionized water if a pH
sensor is not installed). Periodically check the level of the
water to make sure that it does not evaporate.

Method 2: Wet the sponge located in the cap originally included with

the ODO sensor, then install on sensing end of the ODO
sensor. Replace the sponge if it becomes dirty.