Maintenance and storage

pH - pH/ORP sensor storage

The pH - pH/ORP sensors are shipped with their tips in a storage bottle containing KCl. Store the pH - pH/ORP sensors
in the shipping container when not in use.

Figure 71 pH - pH/ORP short-term

storage

pH - pH/ORP sensor short-term

storage (less than 4 weeks)

When in regular field use, the pH-pH/ORP sensors should remain
installed on the bulkhead. Place approximately 0.5 in (1 cm) of any water
(tap or environmental) in the calibration cup (

Figure 71

).

Install the calibration cup onto the bulkhead and firmly tighten to prevent
evaporation.

Figure 72 pH - pH/ORP long-term

storage

pH - pH/ORP sensor long-term storage

Remove the sensor from the bulkhead and insert the sensing end into
the shipping bottle. Install the bottle o-ring and tighten (

Figure 72

).

The shipping bottle contains a 2 molar solution of pH 4 buffer. If this
solution is not available, store the sensor in tap water.

NOTICE: To prevent damage, do not store the pH - pH/ORP

sensors in Zobell solution or DI water.