Calibration

Depth

NOTE: This calibration option is available only if your bulkhead is equipped with a depth sensor. The depth

sensor is located where the cable connects to the bulkhead (

Figure 59 on page 54

).

For the calibration, make sure that the depth sensor is clean and in air, not immersed in any solution. For highest
accuracy, keep the bulkhead still and in one position while calibrating.

NOTE: Cables 10 m and longer are supplied with a weight that can be attached to the sensor guard for sampling

at water depths 10 m and greater.

Figure 53 Calibrate Depth

Depth calibration

1. If applicable, enter the depth offset, altitude, and latitude (

Figure

29 Setup Depth on page 25

).

NOTE: Depth offset allows you to set the depth measurement

to something other than zero. If the depth offset is used,
the depth measurement will be adjusted by the offset
after calibration. Enter the altitude and latitude of your
sampling location to increase the accuracy of your depth
measurement.

2. Push the

Cal

key, then select Depth.

3. Observe the actual measurement readings for stability (white line

on graph shows no significant change for 40 seconds), then select

Accept Calibration (

Figure 53

). “Calibration successful!” will be

displayed in the message area.