Calibration – YSI ProDSS User Manual

Page 40

Advertising
background image

38

Calibration

Figure 48 Calibrate ODO mg/L

ODO mg/L calibration

1. Place the ODO and conductivity/temperature sensor into a water

sample that has been titrated by the Winkler method to determine
the dissolved oxygen concentration in mg/L.

2. Push the

Cal

key, then select ODO. Select DO mg/L.

3. Select Calibration value.
4. Enter the dissolved oxygen concentration of the sample in mg/L.
5. Observe the actual measurement readings for stability (white line

on graph shows no significant change for 40 seconds), then select

Accept Calibration (

Figure 48

). “Calibration successful!” will be

displayed in the message area.

6. Rinse the bulkhead and sensors in clean water then dry.

Figure 49 Calibrate ODO zero point

ODO zero point calibration

1. Place the ODO and Conductivity/Temperature sensors in a

solution of zero DO.

NOTE: A zero DO solution can be made by dissolving

approximately 8-10 grams of sodium sulfite into 500 mL
of tap water. Mix the solution thoroughly. It may take the
solution 60 minutes to be oxygen-free.

2. Push the

Cal

key, then select ODO. Select Zero.

3. Observe the actual measurement readings for stability (white line

on graph shows no significant change for 40 seconds), then select

Accept Calibration (

Figure 49

). “Calibration successful!” will be

displayed in the message area.

4. Thoroughly rinse the bulkhead and sensors in clean water then

dry.

5. Perform a ODO % water-saturated air calibration after performing

a zero point calibration.