Maintenance and storage

Figure 75 ISE short-term storage

ISE sensor short-term storage (less than 4 weeks)

When in regular field use, the ISE sensors should remain installed on the
bulkhead in an environment of water-saturated air. Place approximately
0.5 in (1 cm) of any water (deionized, distilled or environmental) in the
calibration cup (

Figure 75

).

Install the calibration cup onto the bulkhead and firmly tighten to prevent
evaporation.

Figure 76

ISE long-term storage

ISE sensor long-term storage

NOTICE: Do not let the ISE junctions dry out. Junctions that have

been allowed to dry out by improper storage may be
irreparably damaged by dehydration and will require
replacement.

1. Place a small amount of high-calibration solution or tap water in

the storage bottle originally included with the sensor.

2. Remove the sensor from the bulkhead and insert the sensing end

into the shipping bottle.

3. Install the bottle o-ring and tighten (

Figure 76

).

NOTICE: The sensors should not be immersed in water.

NOTICE: Do not store the ISE sensors in conductivity standard, pH

buffer or salt water.

Rehydrating the reference junction

If an ISE sensor has been allowed to dry, soak the sensor for several
hours (preferably overnight) in the sensor’s high-calibration solution. If
the sensor is irreparably damaged, the sensor module must be replaced.