Dissolved oxygen sensor, Maintenance and storage – YSI ProDSS User Manual

Page 59

Advertising
background image

57

Maintenance and storage

Figure 64 Conductivity/Temperature

Long-term storage

Conductivity/temperature sensor long-term storage

The Conductivity/Temperature sensor can be stored dry or wet, installed
on the bulkhead or detached (

Figure 64

).

Dissolved oxygen sensor

ODO sensor caps are warranted for 1 year but have a typical working life of 18 to 24 months. As the ODO sensor
caps ages, large scratches in the paint/dye layer and changes in the dye layer can reduce measurement stability and
response time.

Periodically inspect the sensor cap for damage and large scratches in the paint/dye layer. Replace the cap when
readings become unstable and cleaning the cap and DO recalibration do not remedy the symptoms.

Cleaning the sensor cap

The sensor cap should be kept clean since some types of fouling may consume oxygen which could affect the
dissolved oxygen measurements. To clean the sensor cap, gently wipe away any fouling with a lens cleaning tissue that
has been moistened with water.

NOTICE: Do not use organic solvents to clean the sensor cap. Using an organic solvent to clean the sensor cap

may cause permanent damage to the cap. For example, alcohol will dissolve the outer paint layer and
other organic solvents will likely dissolve the dye in the cap.

ODO sensor cap replacement

The sensor cap should be replaced about once per year but may last longer. It should also be replaced if it is cracked
or damaged. The instruction sheet shipped with the replacement ODO sensor cap includes the calibration coefficients
specific to your sensor cap.

NOTE: Make sure to save the ODO sensor cap instruction sheet in case you need to reload the calibration

coefficients.

NOTE: The replacement ODO sensor cap is shipped in a humidified container and the package should not be

opened until immediately before sensor cap replacement.

Once the sensor cap has been installed on the ODO sensor, it is important to keep the sensor in a 100% humid
environment. If the sensor dries out, refer to the rehydration procedure (

ODO sensor rehydration on page 60

).

Advertising