Operation Theory

33

Figure 4-1: Scan Timing

There are four trigger modes to start the scan acquisition. Refer to
section 4.1 for details. The data transfer mode is discussed in the
following section.

NOTES

The maximum A/D sampling rate is 250 MHz. Therefore,

the minimum setting of SI2_counter is 160 when using

the internal TIMEBASE.

The SI_counter is a 24-bit counter and the SI2_counter is

a 16-bit counter. The maximum scan interval using the in-

ternal Timebase = 2

24

/40 Ms = 0.419 s, and the maxi-

mum sampling interval between two channels using the

internal Timebase = 2

16

/40 Ms = 1.638 ms.

The scan interval may not be smaller than the product of

the data sampling interval and the NumChan_counter

value. The relationship can be represented as:

SI_counter>=SI2_counter * NumChan_counter.

Acquisition_in_progress

Scan_start

AD_conversion

Scan_in_progress
(SSHOUT)(pin8 on CN2)

3 Scans, 4 Samples per scan
(PSC_Counter=3, NumChan_Counter=4)

Sampling Interval t=
SI2_COUNTER/TimeBase

Scan Interval T=
SI_COUNTER/TimeBase

Ch2

Ch3

Ch1

Ch0

Ch2

Ch3

Ch1

Ch0

Ch2

Ch3

Ch1

Ch0

( channel sequences are specified in Channel Gain Queue)