68

Operation Theory

Figure 4-33: High-Hysteresis Analog Trigger Condition

Low-Hysteresis Analog Trigger Condition

Figure 4-34 shows the low-hysteresis analog trigger condition,
the trigger signal is generated when the input analog signal
level is less than the Low_Threshold voltage, and the
High_Threshold voltage determines the hysteresis duration.
Note the High_Threshold setting should be always higher then
the Low_Threshold voltage setting.

Figure 4-34: Low-Hysteresis Analog Trigger Condition