Thresholds 2 – EXFO VDSL2 User Manual
Page 44
 
Auto CQ/Video-DSL Rate Prediction
36
AXS-200/635
Configuring CQ Auto Tests
Thresholds 2
The Thresholds 2 tab allows you to set 
additional threshold values.
Each parameter is described below:
³
Min. VF Long. Balance(dB) 
defines the minimum voice 
frequency threshold longitudinal 
balance level. Specify a value 
between 10 and 80.
³
Max. VF RMS Noise defines the maximum threshold RMS noise energy 
level. Specify a value between -90 and 10 dBm if the ITU standard is 
used. Specify a value between 0 and 100 dBrn if the ANSI standard is 
used.
³
Max. VF Power Influence defines the maximum threshold power 
influence level on the circuit. Specify a value between -90 and 20 dBm 
if the ITU standard is used. Specify a value between 0 and 110 dBrnC if 
the ANSI standard is used.
³
Min. WB Long. Balance(dB) defines the minimum threshold 
wideband longitudinal balance level. Specify a value between 
10 and 80.
³
Attenuation(dB) defines the threshold wideband attenuation level. 
Specify a value between 0 and 80.
To set a threshold value:
1. Press the up/down arrow keys to highlight the desired parameter.
2. Press
to select the value.
3. Press the left arrow key to erase the existing value, and then use the
alphanumeric keypad to enter a value. To cancel the entry, press
.
4. Press
to accept the value.