Auto CQ/Video-DSL Rate Prediction
Copper, VDSL2, ADSL2+, and IP Triple-Play Test Set
49
Configuring Video-DSL Rate Prediction
Impulse Noise Setup
The Imp. Noise Setup tab allows you to configure parameters to measure 
impulse noise on the circuit under test.
Each parameter on the pane is 
described below.
³
Termination defines if the unit 
uses normal test impedance 
termination or high bridging 
impedance when the cable is 
active or terminated by other 
external equipment. For normal 
test termination select 
TERMINATED, otherwise select BRIDGING.
³
Termination Impedance defines the impedance of the dummy load 
connected to the line. Select one of the following: 100 or 135 Ohm.
³
Noise Filter defines the level of wideband noise filtering. The available 
choices are: NONE, 50 kbit, ISDN-E, HDSL-F, ADSL-G, ADSL, ADSL2+, 
VDSL, VDSL2-8, VDSL2-12, VDSL2-17, or VDSL2-30.
Note: VDSL2-x filters are only available if VDSL2WB is enabled in the software
options.
³
Threshold(dBm) is the maximum impulse noise level. Specify a value 
between -50 and 0 (-40 and 0 if any filter applied). Default value is -30.
³
Measurement Time defines the duration of the Impulse Noise test. The 
available choices are: 1, 3, 5, 10, 15, or 60 Minutes, or 24 Hours.