IQS-1700

Solving Common Problems

* * * * * * * is

displayed as power

value.

Channel not

active.

Channel not

selected for

Min. / Max.

acquisition.

Click the Active button for

the appropriate channel in

the Channel tab of the
Settings function tab.

Select the appropriate

channel in Min./Max.
Channel Selection
of the
Instrument function tab.

! ! ! ! ! ! ! is

displayed as power

value.

The signal is unstable

or has a modulated

input.

Select an appropriate manual

measurement range if
Autorange mode is selected.

Questionable

readings displayed.

The detector or

optical connectors

are dirty.

Clean the detector and all optical

connections.

The wavelength

selection is improper.

Switch to the correct wavelength

on all instruments being used.

The offset nulling is

incorrect.

Perform an offset nulling with

protective cap installed.

The optical source is

unstable.

Wait for source to stabilize (at

least 60 minutes).

The correction factor

is incorrect.

Reset the correction factor to

0.000 dB or 1.000 W/W.

Problem

Probable Cause

Recommended Action