Sense[1..n]:correction:factor [:magnitude – EXFO IQS-1700 High Performance Power Meter for IQS-500/600 User Manual

Page 152

Advertising
background image

SCPI Command Reference

144

IQS-1700

Product-Specific Commands—Description

:SENSe[1..n]:CORRection:FACTor

[:MAGNitude]?

Description

This query returns the correction factor. The

value is in W/W units.

Syntax

:SENSe[1..n]:CORRection:FACTor[:MAGNitude]?[

<wsp>MINimum|MAXimum|DEFault]

Parameter(s)

Parameter 1:
The program data syntax for the first parameter is

defined as a <CHARACTER PROGRAM DATA>

element. The allowed <CHARACTER PROGRAM

DATA> elements for this parameter are:

MINimum|MAXimum|DEFault.

MINimum is used to retrieve the instrument's

smallest supported value.

MAXimum is used to retrieve the instrument's

greatest supported value.

DEFault is used to retrieve the instrument's

default value.

Response Syntax

<CorrectionFactor>