A technical specifications, Mportant, High performance power meter – EXFO IQS-1700 High Performance Power Meter for IQS-500/600 User Manual

Page 113: Specifications, Iqs/ohs-1700 series), General specifications

Advertising
background image

High Performance Power Meter

105

A Technical Specifications

I

MPORTANT

The following technical specifications can change without notice.
The information presented in this section is provided as a reference
only. To obtain this product’s most recent technical specifications,
visit the EXFO Web site at www.exfo.com
.

SPECIFICATIONS

a

(IQS/OHS-1700 SERIES)

Optical heads must be operated with the IQS-1710/1720/1740 High-Performance Interface Module.

Model

IQS-1712X/1722X/1742X

IQS-1713/1723/1743

OHS-1713-UH

Number of detectors

1/2/4

1/2/4

1

Detector type

GeX

InGaAs

InGaAs and integrating cavity

Detector size

2 mm detector

1 mm detector

9 mm input aperture

Wavelength range (nm)

800 to 1660

800 to 1700

930 to 1660

Power range (dBm)

b, h

25 to

<50

8 to

<87

37 to

<55

Uncertainty

±(5 % + 10 nW)

e, h

±(5 % + 2 pW)

f, h

±(4 % + 3 nW)

g, h

Polarization-dependent responsivity (dB)

c

N/A

N/A

(0 dBm to

<50 dBm)

±0.008 typ.

Linearity

d

±0.015 dB (10 dBm to

<30 dBm)

±0.015 dB (5 dBm to

<55 dBm)

±0.11 dB (35 dBm to 30 dBm)
±0.05 dB (30 dBm to 5 dBm)
±0.015 dB (5 dBm to −22 dBm)

Power resolution (dB)

h

0.001 (25 dBm to

<20 dBm)

0.001 (8 dBm to

<50 dBm)

0.001 (37 dBm to

<25 dBm)

Wavelength resolution (nm)

0.01

0.01

0.01

Stabilization time (ms)

0.3

0.3

0.3

Sampling rate (sample/s/channel)

up to 5208

up to 5208

up to 5208

Fiber type (μm)

5/125 to 62.5/125
NA

) 0.3

5/125 to 62.5/125
NA

) 0.3

5/125 to 62.5/125
NA

) 0.4 (with BFA-3000)

Notes

a. At 1550 nm (unless otherwise specified), with an FC angled connector and a warmup time

of 20 minutes, followed by an offset nulling.

b. From 18 °C to 28 °C.

c.

At 23 °C ± 3 °C, constant wavelength (1550 nm), constant power and with an FC non-angled
connector.

d. At constant temperature in the 0 °C to 40 °C range; nulling required.

e.

At 23 °C ± 1 °C with an FOA-322 and an FC non-angled connector, between 1000 nm and
1650 nm. Add 1 % to uncertainty below 1000 nm, and 3 % over 1650 nm.

f.

At 23 °C ± 1 °C with an FOA-322 and an FC non-angled connector, between 1000 nm and
1640 nm. Add 1 % to uncertainty below 1000 nm, and 6 % over 1640 nm.

g.

At 23 °C ± 1 °C with an FOA-322 and an FC angled connector, between 1290 nm and 1340 nm,
and between 1420 nm and 1640 nm. Add 2 % to uncertainty below 1000 nm, 1 % between
1370 nm and 1420 nm, and 5 % over 1640 nm. All uncertainties valid on the day of calibration.
Wavelength must not be equal to any water absorption line.

h. Averaging time of 1 s.

i. For optical power of > 35 dBm, maximum operating temperature is 30 °C. With the FOA-396,

maximum operating temperature is 25 °C.

j. From 0 °C to 40 °C.

GENERAL SPECIFICATIONS

IQS-1710/1720/1740
IQS-1713/1723/1743
IQS-1712X/1722X/1742X

OHS-1713-UH

Size (H x W x D)

125 mm x 36 mm x 282 mm
(4

15

/

16

in x 1

7

/

16

in x 11

1

/

8

in)

42 mm x 79 mm x 190 mm
(1

5

/

8

in x 3

1

/

8

in x 7

1

/

2

in)

Weight

0.64 kg (1.4 lb)

0.5 kg (1.1 lb)

Temperature

Operating

i

Storage

0 °C to 40 °C
(32 °F to 104 °F)
<40 °C to 70 °C
(

<40 °F to 158 °F)

0 °C to 40 °C
(32 °F to 104 °F)
<40 °C to 70 °C
(

<40 °F to 158 °F)

Number of ports

1/2/4

1

Relative humidity

j

0 % to 80 %
non-condensing

0 % to 80 %
non-condensing

Remote control

With IQS-600: GPIB (IEEE-488.1, IEEE-488.2), Ethernet and RS-232.

Instrument drivers

LabVIEW™, SCPI commands and COM/DCOM library drivers.

Standard accessories

User guide, one fiber-optic adapter per channel, Certificate of Compliance and Certificate of Calibration.

Advertising