EXFO IQS-1700 High Performance Power Meter for IQS-500/600 User Manual

Page 135

Advertising
background image

SCPI Command Reference

High Performance Power Meter

127

Product-Specific Commands—Description

:MMEMory[1..n]:ACQuisition:DURation:

MMAXimum

Description

This command is used to set the duration of the

Min. Max. acquisition.

Syntax

:MMEMory[1..n]:ACQuisition:DURation:MMAXim

um<wsp><TimeSecond>

Parameter(s)

TimeSecond:
The program data syntax for <TimeSecond> is

defined as a <DECIMAL NUMERIC PROGRAM

DATA> element.

Changes the Min. Max. acquisition duration in

seconds.

Example(s)

MMEM:ACQ:DUR:MMAX 120

See Also

MMEMory:ACQuisition:DURation:MMAXimum?

MMEMory:ACQuisition:DURation:MMAXimum:ST

ATe

MMEMory:ACQuisition:DURation

MMEMory:ACQuisition

INITiate:AUTO

INITiate:EXTRema