Step 5 – calibrate image density, Figure 8, Locate – Datalogic Scanning Matrix 400 User Manual

Page 17

Advertising
background image

RAPID CONFIGURATION

7

1

FOV

Blue Diamond™

in focus

Figure 8 – Focus Function Using Blue Diamonds™

6. Exit the Focus function by pressing the X-PRESS™ push button once. The Blue

Diamonds™ turn off.

7. Replace the lens cover, screwing it tightly to the base.

STEP 5 – CALIBRATE IMAGE DENSITY


In order to function correctly to the fullest extent of its capabilities, Matrix 400™ must acquire
information regarding image density or PPI (pixels per inch). This calibration takes place
through the X-PRESS™ Interface and the Grade A Barcode Test Chart included in the
package. This procedure is necessary for the first time installation, if the lens type is changed
or if the focal distance is changed.

Locate

1. Enter the Focus function by pressing and holding the

X-PRESS™ push button until the Focus LED is on.

2. Release the button to enter the Focus function. The

Blue Diamonds™ turn on.

3. From the Grade A Barcode Test Chart, select the

longest code whose length fits between the two Blue
Diamonds™. Rotate the code 90 degrees and
position the code at the center of the FOV
(equidistant from the Blue Diamonds™).

4. Exit the Focus function by pressing the X-PRESS™

push button once. The Blue Diamonds™ turn off.

READY

green

green

yellow

yellow

red

SETUP

FOCUS

TEST

LEARN

GOOD

TRIGGER

COM

STATUS

Figure 9 – X-PRESS™ Interface:

Locate Function