Kipp&Zonen COMBILOG Data Logger User Manual

Page 103

Advertising
background image

format

length

range of values

integer

2 byte

(dec - 32768)

i

(dec +32767)

Real

4 Byte

(dec – 2

129

)

x

(dec + 2

129

)

Table 11.4 Format settings for transmission in MODBUS-format

Example: The value 50.3094 is to be displayed.

Transmission in ASCII-format:

decimals

field length 6

field length 7

field length 8

0

_ _ _ _ 5 0

_ _ _ _ _ 5 0 _ _ _ _ _ _ 5 0

1

_ _ 5 0 . 3

_ _ _ 5 0 . 3 _ _ _ _ 5 0 . 3

2

_ 5 0 . 3 1

_ _ 5 0 . 3 1 _ _ _ 5 0 . 3 1

3

5 0 . 3 0 9

_ 5 0 . 3 0 9 _ _ 5 0 . 3 0 9

4

E . 3 0 9 4

5 0 . 3 0 9 4 _ 5 0 . 3 0 9 4

5

-

E . 3 0 9 4 0 5 0 . 3 0 9 4 0

6

-

-

E . 3 0 9 4 0 0

Table 11.5 Output formats for transmission in ASCII-format ("_":blank).

Transmission in PROFIBUS and MODBUS-format:

decimals

bool

integer

real

0

-

00 32 (50)

42 49 3C D3 (50 . 3094)

1

-

01 F7 (503)

42 49 3C D3 (50 . 3094)

2

-

13 A6 (5030)

42 49 3C D3 (50 . 3094)

3

-

xx xx (50309)

42 49 3C D3 (50 . 3094)

4

-

xx xx (503094)

42 49 3C D3 (50 . 3094)

5

-

xx xx (5030940)

42 49 3C D3 (50 . 3094)

6

-

xx xx (50309400)

42 49 3C D3 (50 . 3094)

Table 11.6 Output formats for the transmission to PROFIBUS / MOD-

BUS-format (the decimal notation is given in brackets).

Hardware Manual COMBILOG 1022

103