Kipp&Zonen COMBILOG Data Logger User Manual

Page 132

Advertising
background image

Preset Single Register

Description:
By this command a single register can be written.

Request Telegram

ADR

FNR

REGSTA

REGNUM

CRC

6

MSB

LSB

MSB

LSB

MSB LSB

Response Telegram

ADR

FNR

REGADR

DATA

CRC

6

MSB

LSB

MSB

LSB

MSB LSB

ADR ............ module address (hex 00..7F)
FNR ............ function number (hex 06)
REGADR ..... address of the register to be written
DATA …….... data word (hex 0000…FFFF)
CRC ............ check sum
CRC polynomial: u

15

+ u

13

+ 1

CRC start value: hex FFFF

Hardware Manual COMBILOG 1022

132

Advertising