3 resistance measurement – Kipp&Zonen COMBILOG Data Logger User Manual

Page 41

Advertising
background image

5.1.3

Resistance Measurement

Resistance measurement is carried out by means of voltage
measurement at a resistor, measuring the resulting voltage
drop. The constant current required for the resistance meas-
urements is provided by the internal supply of the data logger.

For this purpose the sensor module connects a supply point in-
ternally with the analogue measurement input via a reference
resistor R

o

. The voltage drop U

o

via resistor R

o

is required as a

reference for further signal processing by the module. The res-
istance value of the sensor can be calculated from the input
signals U

i

as a multiple of the reference resistor R

o

. Measuring

range is between 0 and 20 k

.

Note: The data logger COMBILOG 1022 supports resistance
measurement in 2-, 3- and 4-wire technique. With resistance
measurement in 2-wire technique the supply lines cause an ad-
ditional voltage drop, thus distorting the measuring result and
influencing the measuring accuracy. Therefore it is necessary
to pay attention especially with resistance measurement in 2-
wire-technique. Wires with impedance as low as possible
should be used. Make sure that the leads are well connected to
the data logger and the sensor. With resistance measurement
using 3-wire technique the potential on the supply lines will be
subtracted by software. Therefore 2 measurements are neces-
sary, resulting in double measuring time. With resistance
measurement in 4-wire technique the drop of potential is picked
up directly at the sensor, so that the measuring results are not
influenced by the supply lines.

Hardware Manual COMBILOG 1022

41

Advertising