Kipp&Zonen COMBILOG Data Logger User Manual

Page 140

Advertising
background image

Read data from external channel

Register

Type Contents

Range

2000

ro

Channel 1 real High Word

0 ... 65535

2001

ro

Channel 1 real Low Word

0 ... 65535

2002

ro

Channel 2 real High Word

0 ... 65535

2003

ro

Channel 2 real Low Word

0 ... 65535

:

:

:

:

2076

ro

Channel 60 real High Word

0 ... 65535

2077

ro

Channel 60 real Low Word

0 ... 65535

Attention: The low word and the high word of a channel must
always be read or written simultaneously.

Read data logger

Register

Type Contents

Range

3000

ro

Number of records

0 ... 65535

3001

ro

Number of values in a record 0 ... 65535

3002

ro

Value 1 real High Word

0 ... 65535

3003

ro

Value 1 real Low Word

0 ... 65535

3004

ro

Value 2 real High Word

0 ... 65535

3005

ro

Value 2 real Low Word

0 ... 65535

:

:

:

:

30B8

ro

Value 92 real High Word

0 ... 65535

30B9

ro

Value 92 real Low Word

0 ... 65535

Note: The oldest record will be load to the register 3002 to
30B9 while reading register 3001. The number of values de-
pends on the configuration.

Reserved for download functions

Register
FD00..FFFF

Hardware Manual COMBILOG 1022