Chapter 2: Prerequisites

Product Service 1-800-522-7658

6


Labconco offers accessory remote blowers and transition adapters listed in Chapter 7.
Labconco also offers accessory FilterMate Portable Exhausters listed in Chapter 7.
Contact Labconco for blower sizing assistance.

If the enclosure is connected directly to a house exhaust system,
an adjustable damper (or valve) must be installed to control the
airflow properly. This is equally important when a house exhaust
system is controlling multiple enclosures. See Chapter 7 for
accessory Adjustable Damper ordering information.

Electrical Requirements

Standard duplex electrical receptacles should be nearby for connecting the
enclosure or other equipment. The enclosures for remote blower exhaust include
iris pass-throughs to allow electrical cords through the back of the enclosure
without leaving a large hole for contaminants to escape.

The remote blower for the Protector Work Station may be connected to the blower
switch on the enclosure. There is an outlet on the back of the upper light housing
for this specific purpose. We recommend a maximum amperage of 6 amps for this
circuit to the remote blower. Please refer to the wiring diagrams in Appendix C.
Connect the blower wires to the remote blower per local electrical codes.

Space Requirements

The dimensions for the different models are shown in Appendix B: Dimensions.

!