Chapter 1: Introduction
Product Service 1-800-522-7658
2
No.
Use Protector Work
Station Model No’s.
Filter
Blower
Application
Exhaust
1.
3930000 (115V)
3930020 (230V)
Carbon Filter*
Organic Vapor,
Formaldehyde or
Ammonia
Built-In
Integral
Motorized
Impeller
If ducting to the outside is not
feasible, the filters keep low
level concentrations below
OSHA recommended time
weighted averages, restoring
clean air to the laboratory.
To the room
2.
3930001 (115V)
3930021 (230V)
None
Remote
Blower or
House Exhaust
Best choice for high level
concentrations or when chemical
concentrations are unknown or
unpredictable
To the outside
3.
3930001 (115V)
3930021 (230V)
FilterMate
Carbon Filter*
and/or HEPA
Filter for
particulates
FilterMate
Portable
Exhauster
Best choice when particulate
removal is required in addition
to carbon filtration and ducting
to the outside is not feasible
To the room
* It is recommended that a Labconco product specialist review the chemical
application to determine if it is suitable. Consult Chapter 5 and Chapter 6.
Protector Work Station 3930001 is
shown on Work Surface with Sink
6942003 and Telescoping Base
Stand 3746702. Blower, ductwork,
work surface and supporting base
must be ordered separately.