Labconco Protector Work Stations 3930021 User Manual

Page 80

Advertising
background image

Product Service 1-800-522-7658

76







Appendix D:

Quick Chart for the Protector

Work Stations









*excludes remote Protector Work Station, since it has no integral blower.


To determine the actual face velocity at the sash opening, airflow velocity readings
will need to be taken. This should be done across the sash opening of the
enclosure in accordance with the Industrial Ventilation Manual. (See Appendix E)
The “average face velocity” is achieved by taking readings in two rows across the
enclosure with the readings 6" from the ends and evenly spaced every 12"; the first
row is 3" down from the upper sash foil and the second row is 3" up from the work
surface. A total of eight readings are taken for the 4' Protector Work Station and
then averaged.

Model Size

4'

Sash Height from Work Surface (inches)

8.00

Total Open Area with Bypass (sq. ft.)

2.50

Exhaust Airflow Volume at 80fpm (CFM)

200

Alarm Airflow Volume at 60 fpm (CFM)

150

*Motor Horsepower (HP)

0.30

*Capacitor (MFD, Volts)

15MFD, 115V

4MFD, 230V

Fluorescent Lights

F25T8

Advertising