Chapter 3: Getting Started

Product Service 1-800-522-7658

11

Connecting to the Exhaust System

(Required on Protector Work Station for

Remote Exhaust)

ATTENTION: The weight of the exhaust ductwork system
must be supported independently of the enclosure
superstructure or damage may occur.


The exhaust system should be installed by a qualified HVAC
contractor
.


The exhaust connection on the remote Protector Work Station Enclosure has been
designed to accept a 2" x 10" (5.1cm x 25.4cm) nominal transition adapter. See
Chapter 7 for ordering accessory Transition Adapters. Labconco manufactures
transition adapters for either top or bottom exhaust and for both 5" dia. hose or 6"
dia. duct. Review Chapter 2 for remote blower and FilterMate Portable Exhauster
exhaust prerequisites and review Chapter 7 for ordering blower exhaust equipment.
For your convenience several exhaust options are shown in Figures 3-2 and 3-3.
Consult Labconco Customer Service should you require help sizing your blower
for the exhaust volume and system static pressure loss.

To ensure compatibility, the selected exhaust duct material
should match the enclosure, procedures and chemical
applications.

!

!