Chapter 6: Maintaining Your Enclosure

Product Service 1-800-522-7658

47

9. Be sure to enter “DONE” after successfully completing the low and high

calibration set points.

10. Once calibration is completed, go to “RUN” and hit “ENTER”. The value

should read close to the high calibration set point.

11. To lower the face velocity to the operating point, simply alter the exhaust

volume with the speed control on the FilterMate or exhaust damper. Then
recheck the face velocity with an anemometer to confirm the display on the
digital airflow monitor.

Digital 1000 Alternate Calibration Procedure -

Constant Volume Conditions

1. To successfully calibrate, it will be necessary to change the face velocity by

opening and closing the enclosure’s sash. The airflow monitor is factory set
to be calibrated with a difference of at least 50 fpm and can be changed by
adjusting the “lower/higher air sample difference”. The in flow face
velocity speeds provided in the chart below are suggested to successfully
calibrate the Digital 1000.

2. Before proceeding with calibration, it will first be necessary to configure

the airflow monitor. Go to the setup and then CAL CONFIG MENU and
adjust the “lower/higher air sample difference” to 10 fpm. This will allow
you to successfully calibrate with minimum difference values of 10 fpm.

3. While in CAL CONFIG MENU, change the “sensor difference” from 10%

to 3%.

4. While in CAL CONFIG MENU, adjust the red low alarm to the desired

setting (See the chart below for range and suggested settings). Then adjust
the yellow “CAUTION or WARNING” to the desired setting (See the chart
below for range and suggested settings). Then adjust the “CAUTION or
WARNING” air reset to 3 fpm. This sets the alarm condition.

CAUTION WARNING

“YELLOW LED” Setting

LOW ALARM

“RED LED”

Setting

Low Calibration

Set Point

Sash Open

(fpm)

High Calibration

Set Point

or Enclosure

Operating

Inflow Speed

Sash Closed

Range

user defined

(fpm)

Suggested

Set Point

(fpm)

Range

user defined

(fpm)

* Suggested

Set Point

(fpm)

8" Sash

Height

10" Sash

Height

43 – 53

51

40 – 50

48

27

23

60 fpm

63 – 73

67

60 – 70

64

37

30

80 fpm

83 - 93

83

80 - 90

80

47

37

100 fpm

*

Because of airflow fluctuations in a typical laboratory environment Labconco suggests

setting the “RED LED” low alarm set point to 20% below the enclosure’s operating speed.