Labconco Protector Work Stations 3930021 User Manual

Page 53

Advertising
background image

Chapter 6: Maintaining Your Enclosure

Product Service 1-800-522-7658

49

Note: Enter Button stores information and +/- Buttons allow for scrolling.

1. Push the “ENTER” Button on the face of the alarm until the “SET UP”

Menu is displayed.

2. Scroll to “SET UP” and hit “ENTER”.
3. The PASSWORD MENU displays (The Password is 0000). Press the

ENTER button repeatedly until the CAL CONFIG MENU is displayed.

4. In the CAL CONFIG MENU set the following:

CALIBRATION - CONFIGURATION MENU

SETTINGS

DISPLAY UNITS

FPM


LOW AIR ALARM “RED LED”

* 48 FPM (60 fpm Operating inflow)
* 64 FPM (80 fpm Operating inflow)
* 80 FPM (100 fpm Operating inflow)

LOW AIR CUTOFF

OFF


WARNING AIR ALARM “YELLOW LED”

* 51 FPM (60 fpm Operating inflow)
* 67 FPM (80 fpm Operating inflow)
* 83 FPM (100 fpm Operating inflow)

WARNING AIR RESET

3 FPM

HIGH AIR ALARM

OFF

LOWER AIR SAMPLE FLUCTUATIONS

OFF

HIGHER AIR FLUCTUATIONS

3%

LOWER / HIGHER AIR SAMPLE DIFFERENCE

3%

WARN TO ALARM AIR TIME

10 SECONDS

ALARM TO WARN AIR TIME

3 SECONDS

SHOW AIR FLOW

ON

SHOW TIME LINE OFF = DISPLAYS BAR
GRAPH

OFF

AUDIBLE ALARM

ENABLED

SENSOR DIFFERENCE

2%

SENSITIVITY

80%

* (Suggested Air Alarm Settings) Refer to the chart on the previous page for a
range of air velocity settings that may be used.

5. To complete the CAL CONFIG, be sure to enter “DONE”. You are

returned to the Main Menu.