Chapter 6: Maintaining Your Enclosure
Product Service 1-800-522-7658
49
Note: Enter Button stores information and +/- Buttons allow for scrolling.
1. Push the “ENTER” Button on the face of the alarm until the “SET UP”
Menu is displayed.
2. Scroll to “SET UP” and hit “ENTER”.
3. The PASSWORD MENU displays (The Password is 0000). Press the
ENTER button repeatedly until the CAL CONFIG MENU is displayed.
4. In the CAL CONFIG MENU set the following:
CALIBRATION - CONFIGURATION MENU
SETTINGS
DISPLAY UNITS
FPM
LOW AIR ALARM “RED LED”
* 48 FPM (60 fpm Operating inflow)
* 64 FPM (80 fpm Operating inflow)
* 80 FPM (100 fpm Operating inflow)
LOW AIR CUTOFF
OFF
WARNING AIR ALARM “YELLOW LED”
* 51 FPM (60 fpm Operating inflow)
* 67 FPM (80 fpm Operating inflow)
* 83 FPM (100 fpm Operating inflow)
WARNING AIR RESET
3 FPM
HIGH AIR ALARM
OFF
LOWER AIR SAMPLE FLUCTUATIONS
OFF
HIGHER AIR FLUCTUATIONS
3%
LOWER / HIGHER AIR SAMPLE DIFFERENCE
3%
WARN TO ALARM AIR TIME
10 SECONDS
ALARM TO WARN AIR TIME
3 SECONDS
SHOW AIR FLOW
ON
SHOW TIME LINE OFF = DISPLAYS BAR
GRAPH
OFF
AUDIBLE ALARM
ENABLED
SENSOR DIFFERENCE
2%
SENSITIVITY
80%
* (Suggested Air Alarm Settings) Refer to the chart on the previous page for a
range of air velocity settings that may be used.
5. To complete the CAL CONFIG, be sure to enter “DONE”. You are
returned to the Main Menu.