Chapter 6: Maintaining Your Enclosure

Product Service 1-800-522-7658

43

Guardian™ Digital Kit No 3908800 or

3908801

Guardian Digital 1000 Operation

The Guardian Digital Airflow Monitor consists of the airflow sensor, the Alarm Unit
and the 15 VDC power supply. For 115V operation the alarm unit is powered by
plugging the power supply into the factory-prepared digital airflow monitor socket.
For 230V operation, the Alarm Unit is powered by plugging the power supply into a
building outlet. The alarm has “Enter”, “+”, and “-” buttons to program the monitor.
There is also a green LED “SAFE”, yellow LED “CAUTION”, and red LED
“LOW” with audible alarm for airflow conditions. The audible alarm can be
permanently muted if desired. The Guardian Digital 1000 Airflow Monitor displays
a face velocity value, provides an RS232 communications port to a PC or building
computer system, can be configured for external input connections such as night
setback or external alarm and provides up to three output relays that can be
configured. For complete detailed information, please refer to the separate Labconco
1000 Alarm User’s Manual provided with the enclosure.

Guardian Digital 1000 Installation Procedure

1. The enclosure comes prepared to accept the Guardian Digital airflow

monitor system.

2. First remove the large 1.19" dia. gray hole plug. See Figure 6-6. See

Figure 6-7 only to reference internal assembly of the airflow monitor.
Locate the elbow, locknut, and washer and install it in the 1.19" dia. hole
per Figure 6-6 and Figure 6-7. The enclosure baffle pivots down to install
the elbow, washer and locknut.

3. Cut the 1" hose supplied with the kit to 10.5" approximate length and

install it between the airflow sensor and the elbow.

4. Secure the Guardian Digital alarm to the enclosure with double stick tape

as shown in Figure 6-9. The airway passage between the alarm module and
the enclosure is now complete.

5. Locate the metal hose cover and install with double stick tape per Figure 6-8.
6. Locate the power supply transformer. One end should already be

connected to the two-pin connector labeled 15 VDC on the back of the
alarm module and through the strain relief bushing. If disconnected, then
reconnect to power the airflow monitor. Plug the 115V power supply into
a standard 115V duplex receptacle, the back of the accessory FilterMate
portable exhauster or the back of the accessory light. For 230V, plug into a
standard receptacle with your specific outlet plug. (It is recommended
that the airflow monitor be connected directly to the FilterMate
switched auxiliary outlet so the airflow monitor is powered at the same
time.)

7. Installation is now complete.