Labconco Protector Work Stations 3930021 User Manual

Page 39

Advertising
background image

Chapter 6: Maintaining Your Enclosure

Product Service 1-800-522-7658

35


Monthly (or more often as required)

Determine the actual face velocity through the sash opening of the

enclosure where the average reading should be at the specified velocity.
(Use calibrated thermal anemometer or other approved apparatus). Airflow
alarms are recommended for constant monitoring.

The enclosure rear baffle should be checked for any blockage to ensure that

the enclosure is maintaining proper airflow.

All weekly activities.

While the enclosure is filled with the contaminant, test filter condition on

carbon filters using the appropriate gas detector tube at intervals of 20% of
the total estimated time. The exception to the 20% recommendation is
formaldehyde or any carcinogen or suspected carcinogen. These hazardous
chemicals must be checked at least every 10% of the total estimated time.
Gas detector tubes for the specific chemicals that are being used in the
enclosure can be obtained from your laboratory supply dealer.

Replace carbon filters when chemical breakthrough is indicated by odor,

time, detector tube, or for some chemicals, analytical instrumentation. See
“Replacing Carbon Filters” section of this manual in Chapter 6.


Annually

Replace the fluorescent lamps.

Have the enclosure speed and carbon filters (if equipped) validated by a

qualified certification technician.

All monthly activities.

Setting the Inflow Face Velocity with the

Speed Control Adjustment

1. The speed control is located behind the switched control panel. See Figure

6-1.

2. Adjust the speed control with a small Phillips screwdriver by turning the

screw counterclockwise to increase blower speed or clockwise to decrease
the blower speed. The speed control is very sensitive, so proceed with
caution.

3. Measure the inflow velocity per the averaging technique outlined in

Chapter 3 and adjust the speed control slowly for the desired speed. Allow
the speed to stabilize and re-measure the inflow velocity to confirm.


CARBON

Advertising