Chapter 6: Maintaining Your Enclosure

Product Service 1-800-522-7658

38

Guardian 500 Installation Procedure

1. The enclosure comes prepared to except the Guardian™ 500 Airflow

Monitor.

2. First remove the large 1.19" dia. gray hole plug. See Figure 6-3. See Figure

6-4 only to reference internal assembly of the airflow monitor. Locate the
elbow, locknut, and washer and install it in the 1.19" dia. hole per Figure 6-3
and Figure 6-4. The enclosure baffle pivots down to install the elbow, washer
and locknut.

3. Cut the 1" hose supplied with the kit to 15.5" approximate length and install

it between the airflow sensor and the elbow.

4. Secure the Guardian 500 Airflow Monitor to the enclosure with double stick

tape as shown in Figure 6-5. The airway passage between the alarm module
and the enclosure is now complete.

5. Locate the metal hose cover and install with double stick tape per Figure 6-5.

6. Locate the power supply transformer. One end should already be connected

to the two-pin connector labeled 15 VDC on the back of the alarm module
and through the strain relief bushing. If disconnected, then reconnect to
power the airflow monitor. Plug the 115V power supply into a standard
115V duplex receptacle, the back of the accessory FilterMate portable
exhauster or the back of the accessory light. For 230V, plug into a standard
receptacle with your specific outlet plug. (It is recommended that the
airflow monitor be connected directly to the FilterMate switched
auxiliary outlet so the airflow monitor is powered at the same time.)

7. Installation is now complete.